118 . Clarus 590 GC User’s Guide
CAUTION
ATTENTION
To ensure proper auto-ignite FID operation, make sure your
column is positioned in the FID 3.74 inches (95 mm) from the
back of the column nut.
Flame ignition may not occur if your carrier gas flow is 60
mL/min or higher.
Pour assurer un fonctionnement correct de l'auto-inflammation
FID, assurez-vous que votre colonne est positionnée dans le
FID à 2,74 pouces (95 mm) à l'arrière de l'écrou de la colonne.
La flamme peut ne pas s'allumer automatiquement si le débit
de votre gaz porteur est de 60 mL / min ou plus.
1.
Touch the
FID (FIDW)
icon from the System Status screen.
The FID tab page appears:
NOTE: For proper ignition to occur, the FID must be heated to at least 100
°
C.
2.
Touch the temp field
and use the up and down arrow or keypad
buttons to enter a setpoint of at least 100
°
C, or enter a value 50
°
C higher than the highest
temperature of your oven program.
3.
Wait for the FID temperature to become READY. When the temperature is ready a green light
appears next to the temperature display
.
4.
Touch the
Ignite
button
.
(Note the signal level in mV before the flame ignites.)
5.
Upon pressing the Ignite button, the air flow turns off for 15 seconds,
Flame Ignition 1
appears
in the status bar, and the glow plug begins to glow.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...