82 . Clarus 590 GC User’s Guide
Now you can only pressure program your capillary injector. If you wish to program the Capillary
control for more than just pressure skip to step 9.
4.
Under
Flow Offset
, select either the
Auto
or
Fixed
.
The
Auto
mode is the preferred mode of operation. If you want to shorten the overall run time at
the expense of some area precision, then use the
Fixed
mode.
We recommend that if you are operating with a split vent flow of 50 mL/min or less, select the
Auto
mode.
Before each injection in the
Auto
mode, the system briefly closes the split vent and measures the
total flow. With the split vent off this flow is the column flow plus the septum purge flow. The
septum purge flow is maintained due to a mechanical restrictor and the split vent flow and split
ratio are maintained for each injection.
In the
Fixed
mode, you enter an offset value which is equal to the septum purge flow plus the
column flow at the time of injection. Since the system controls the total flow (equal to the split
vent plus septum purge plus column flow), the septum purge is fixed, and the column flow is
controlled, the difference between these two flows and the offset entered will go out the vent flow.
This will be consistent for each run.
5.
If
Fixed
is selected use the up and down arrow or keypad
buttons to enter an offset
value in mL/min.
6.
Press
OK
.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...