![PerkinElmer CLARUS 590 GC User Manual Download Page 226](http://html1.mh-extra.com/html/perkinelmer/clarus-590-gc/clarus-590-gc_user-manual_1530196226.webp)
226 . Clarus 590 GC User’s Guide
External Configuration
Run/ AutoSampler
Pre wash
is number of times the syringe is filled with the sample and emptied to a waste vial
before a sample is acquired for injection. The default is 2. Permissible entries are 0 to 15.
Pump
is the number of times the syringe is filled with the sample before it is acquired for
injection. The default is 6. Permissible entries are 0 to 15.
Post-inject
is the number of times the syringe if filled with solvent and emptied to waste after
sample injection. The first half of these washes is taken from solvent vial 1 or 3. The second half
from solvent vial 2 or 4.The default is 2. Permissible entries are 0 to 15.
Injection Mode
Fast
– 30 millesectonds. Not available when a 50.0 syringe is installed.
Normal
injection takes about 1 second.
Slow
takes about 7.5 second.
Auto resume
provides option for autosamplers program to automatically resume where it left
off should a power failure occur.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...