84 . Clarus 590 GC User’s Guide
10.
Use the up and down arrow or keypad buttons to enter an
Internal
Diameter
which is
measured in micrometers (not millimeters).
11.
Then view the calculated linear gas velocity from selecting
Utilities
from the
Tools
menu and
selecting the
Column Flow Calc.
The Column Carrier Gas Calculator screen appears:
NOTE:
If you incorrectly enter a column length or internal diameter, the resulting flow and pressure
may be incorrect. For example, a 0.25 mm column may actually be 0.22 mm To verify that you
have entered the correct values, you can measure the linear gas velocity by injecting methane
and calculating the linear gas velocity
column length (cm)
Retention Time of Methane (sec)
12.
Return back to the
Configure Injector
screen and select either a
Flow
or
Ratio Split
Mode
.
•
If you select
Flow
, you will enter a split flow in your method. The software controls and
calculates the split ratio.
•
If you select
Ratio
, you enter the split ratio in your method. The software controls the ratio
and calculates the split flow.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...