194 . Clarus 590 GC User’s Guide
Matching Detector Signals w ith Your Integrator
The following procedure should be followed if you are operating your detector near the minimal detectable
quantity level. This will ensure that the integrator is not filtering the signal coming from the Clarus GC. The
purpose of attenuation when an integrator is attached and configured is to assure that the GC output is
above the noise level of the integrator.
Normally, attenuation for an integrator need only be performed once. To obtain the most satisfactory
results, it is recommended that attenuation be carried out under actual operating conditions prior to
running your first sample.
The greater the attenuation, the smaller the degree of detector signal amplification. Attenuation of an
integrator can be thought of as a two-step process:
1.
Observe the integrator noise at an attenuation of 64:
2.
Gradually decrease the attenuation (for example, to 16 or until the noise level increases):
A table listing attenuation vs. the maximum detector signal producing a 1-V output to an integrator
is provided in the Clarus GC Hardware Guide, Maintenance chapter, section “Practical Hints.”
To change an attenuation value:
1.
From the System Status screen touch the icon correcponding to your detector.
2.
In the tab page touch the
Atten
field
and use drop down menu to change the value.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...