Glossary of Clarus GC Terms . 19
Glossary of Chromatographic Terms
Adsorption
– A process that occurs at the surface of a liquid or solid as a result of the attractive forces
between the adsorbent and the solute. These forces may be physical or weakly chemical.
Analysis
– The complete investigation of a sample by gas chromatographic separation including
identification of the sample components and quantitative measurements.
Anode
– The negatively charged electrode in any electrical circuit to which charged particles and ions are
attracted.
Band Broadening
– A process that occurs in the GC whereby the peak width for a component increases
the longer the component travels through the column.
Baseline
– The detector signal to a recorder or integrator when only the carrier gas is passing through
the detector.
Baseline drift
– Any regular change occurring in the baseline signal from the detector, usually resulting
from column temperature and/or gas flow changes.
Blank run
– A run without the sample being injected.
Bleed
– The evaporation of the stationary phase from a column.
Capillary column
(wall coated open tubular column) – A small-internal-diameter column whose inside
wall is coated with a liquid phase.
Carrier gas
– The mobile phase of the separation system. An inert gas which transports the sample
from the injector through the column to the detector. This gas is usually helium, hydrogen, or nitrogen.
Column conditioning
– A process for producing a stable column by heating the column with carrier gas
flowing to remove volatile impurities from the stationary phase.
Detectors
– Hardware that responds to sample components producing an electrical signal that can be
measured to quantitate the amount of each component present.
Flow programming
– The stored PPC program is interpreted as real column mass flow settings or
capillary column flow settings from which the inlet pressure is calculated. This is dependent on the
capillary control mode and installed pressure or split control option.
Flow rate
– The mass flow of carrier gas or detector gas in milliliters per minute.
Ghost peaks
– Peaks that are not due to sample components, for example, peaks produced by carrier
gas impurities, septum, or components from previous analyses..
Injection port
– The hardware through which the sample is introduced to the column by injection.
Linearity
– Quantitatively all detectors will produce a linear response with respect to solute
concentration over a defined range, for example, the Linear Range.
Liquid phase
– The material in the column that causes the components to separate because of
partitioning of the components between the mobile phase (carrier gas) and the stationary phase (liquid
phase).
Low est limit of detection
– The smallest amount of sample that can be detected by the detector being
used. Usually defined as any signal that is as great as two times the noise level. Also referred to as
Minimal Detectable Quantity (MDQ).
*
Reference: Denney, R.C.
A Dictionary of Chromatography.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...