FID Overview . 113
FID Overview
Flame Ionization Detectors (FID) are mass sensitive and used for the destructive analysis of organic
compounds.
As the sample enters the detector, it is mixed with hydrogen and then burned in air, thereby generating
ions. The ions are collected and measured by the detector, with their concentration proportional to the
amount of compound present. Ion collection is enhanced by a polarized electric field that is created by
applying a voltage (
+50
V) to the jet tip. The detector sensitivity is affected primarily by the hydrogen flow.
The FID has
flameout
detection and an
auto-ignite
feature. Flameout detection occurs 0.5 minutes into
the
PRE-RUN
time and it measures the user-entered baseline threshold level in mV. If the threshold level
is exceeded, the flame is considered lit, otherwise the Clarus GC detects that the flame is out and auto-
ignite automatically lights the flame before the start of a run. You can also manually light the flame at any
time by pressing the Ignite button
.
A cross section view of a FID is shown below.
Setting Up a PPC FID
To set up a FID that uses PPC modules, first read the FID setup summary, then follow these procedures.
FID Setup Summary
To set up the PPC version of the FID:
•
Set the air and hydrogen flows.
•
Automatically ignite the flame.
•
Manually ignite the flame.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...