![PerkinElmer CLARUS 590 GC User Manual Download Page 111](http://html1.mh-extra.com/html/perkinelmer/clarus-590-gc/clarus-590-gc_user-manual_1530196111.webp)
Overview . 111
Overview
This section contains procedures for setting up the following detectors:
FID (flame ionization)
PID (photoionization)
ECD (electron capture)
NPD (nitrogen phosphorus)
TCD (thermal conductivity)
FPD (flame photometric)
The instructions provided in this chapter describe how to set detector gases that are controlled by PPC
modules.
CAUTION
ATTENTION
In a Clarus GC equipped with PPC modules, the modules are calibrated
at the factory with the corresponding gas; however, there are
exceptions. The PPC module that controls the:
•
ECD is calibrated with nitrogen.
•
TCD is calibrated with helium.
•
PID is calibrated with helium.
If you intend to use a gas that is different from the gas used to
calibrate the PPC module, you must recalibrate the PPC module with
the gas.
Dans un Clarus GC équipé de modules PPC, les modules sont étalonnés
en usine avec le gaz correspondant; Cependant, il existe des
exceptions. Le module PPC qui contrôle le:
•
L'ECD est calibré avec de l'azote.
•
Le TCD est calibré avec de l'hélium.
•
Le PID est calibré avec de l'hélium.
Si vous avez l'intention d'utiliser un gaz différent du gaz utilisé pour
étalonner le module PPC, vous devez recalibrer le module PPC avec le
gaz.
The instructional procedures in this chapter are written with the assumption that:
You know how to measure gas flows using a flowmeter with the Clarus GC’s built-in stopwatch. If not,
prior to continuing, please read "Using the Built-in Stopwatch" in Chapter 4, "Before You Install a Column."
•
All heaters are turned
OFF
and cool to the touch.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...