![PerkinElmer CLARUS 590 GC User Manual Download Page 180](http://html1.mh-extra.com/html/perkinelmer/clarus-590-gc/clarus-590-gc_user-manual_1530196180.webp)
180 . Clarus 590 GC User’s Guide
3.
To turn on the photomultiplier tube touch the
button in FPD tab screen then touch
Configuration
from the drop down menu . This will bring you to the FPD Configuration screen
page.
4.
In the PMT Voltage field set the percentage (see note below) by using the plus minus buttons. This
will turn on the photomultiplier tube.
NOTE: The voltage to the photomultiplier tube has been set in the factory to approximately 70%. You can change
this setting from the Clarus GC touchscreen. You may want to change this setting as the photomultiplier
tube ages and the signal diminishes. The signal-to-noise ratio may vary with each photomultiplier tube;
therefore, increasing the voltage may not increase the signal-to-noise ratio.
After installing a replacement photomultiplier tube, the noise should be about 2% when the Clarus GC
integrator output is set to attenuation 8 (x8) and the linearizer off. See the
Clarus 590 GC Hardware Guide
"Maintenance” chapter, for this procedure.
NOTE: The photomultiplier tube can be damaged if exposed to light while voltage is ON. Never remove the
detector cap or Photomultiplier tube WITH the voltage on.
Ignite the Flame
1.
Open the electronics cover to access the ignitor assembly.
2.
From the system status screen touch the
FPD
icon.
3.
The FPD tab page is now visible.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...