FID Overview . 115
The default setting is 200 milliseconds. To improve the signal-to-noise, change the setting to 800
milliseconds. This improves the noise but broadens the peak shapes. If you want a faster detector
response, change the setting to 50 milliseconds; however, the noise level may increase.
5.
Touch the
Set
Point
dialog box for the FID
Flameout
; and then use the up and down arrow or
keypad
buttons to enter a flameout detection level in mV (for example 0.1 mV).
NOTE: You must set a Flameout value in millivolts (mV) for the FID auto-ignition feature to work.
For example, 0.3 mV.
Flameout detection occurs 0.5 minutes into the PRE-RUN time and measures the user-entered
baseline threshold level in mV. In this example 0.3 mV are used. If the 0.3 mV threshold level is
exceeded, the flame is considered lit; otherwise, the auto-ignite sequence begins to automatically
ignite the flame.
6.
Touch the
Set Point
dialog box for the
Air
flow and use the plus and minus buttons to set the FID
Air flow to
450
mL/min.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...