2-8
Alarm terminal position
Enclosed in parentheses are the relay numbers assigned to each option terminal.
Option terminal
Terminal arrangement (only the position of the alarm output terminals is
indicated.)
When the relay is not energized, NC is closed and NO is open. C is the common
terminal.
/A1, /A1/F1
/A2, /A2/F1
/A3, /A3/R1
Contact Specifications
Item
Specification
Output type
Relay transfer contact (energize/de-energize switchable)
Output capacity
250 VAC (50/60 Hz), 3 A
250 VDC, 0.1 A (resistive load)
Dielectric strength
1500 VAC (50/60 Hz) for one minute between output terminals and
earth ground terminal
For details related to the switching of energized/de-energized state of the alarm relays,
see section 6.4.
2.4 Alarm Output Wiring (/A1 to /A3, Option)
Summary of Contents for RD-MV100 Series
Page 40: ...1 22 This page intentionally left blank...
Page 41: ...1 23 Overview of Functions 1 This page intentionally left blank...
Page 139: ...7 5 Setting and Operating the Display 7 This page intentionally left blank...
Page 140: ...7 6 This page intentionally left blank...
Page 215: ...10 15 Other Functions 10 This page intentionally left blank...
Page 216: ...10 16 This page intentionally left blank...
Page 217: ...10 17 Other Functions 10 This page intentionally left blank...
Page 218: ...10 18 This page intentionally left blank...
Page 219: ...10 19 Other Functions 10 This page intentionally left blank...