11-23
Computation/Report Function (Option)
11
For example, if the scan interval is 2 s, and the input value is 100 m
3
/min, a simple
summation would add 100 every 2 s resulting in 3000 after one minute. However, if the
sum unit is set to /min, then 2s/60s is multiplied every scan interval before the value is
added giving a result that is close to the actual input value. The scan interval unit is in
seconds.
OFF
∑
(measured value)
/S
∑
(measured value)
×
scan interval
/MIN
∑
(measured value)
×
scan interval/60
/HOUR
∑
(measured value)
×
scan interval/3600
/DAY
∑
(measured value)
×
scan interval/86400
For the setting procedure, see section 11.13.
If power is lost while the report function is in progress
If a power disruption occurred while the report function was in progress, the report
function will resume after the power is restored. The exact operation will vary depending
on whether the power is restored before or after the scheduled time to create a report.
Time of Recovery
Report Operation
After the time to create the
Report data are created immediately after power is restored. The
report
measured/computed data up to the time of the power disruption are
used. For the next scheduled report, data after the power recovery
are used.
Before the time to create
After power is restored, report data are created at the time of the
the report
next normally scheduled report. All measured/computed data
excluding the power disruption period are used.
How the measured/computed data are handled when there is an abnormality in the
data
Type of Abnormal Data
Report Data
Average Value Maximum/Minimum Value
Integrated Value
Positive overflow
not used
used
not used
Negative overflow
not used
used
not used
Error
not used
not used
not used
Displaying the Report
Report data display
The reports can be displayed from the [Information] - [Memory Summary] menu in the
operation mode. See section 4.5.
Status display
If the data of a measurement or a computation channel enter any of the conditions listed
below within the relevant time period (one hour for hourly reports and one day for daily
reports, for example), status “E” or “O” is output to the report. In addition, if a power
failure occurred or the time was changed, status “P” and “C” are output to the report,
respectively. For the displayed position, see the report data output example in section
4.5.
11.11 Overview of the Report Function
Summary of Contents for RD-MV100 Series
Page 40: ...1 22 This page intentionally left blank...
Page 41: ...1 23 Overview of Functions 1 This page intentionally left blank...
Page 139: ...7 5 Setting and Operating the Display 7 This page intentionally left blank...
Page 140: ...7 6 This page intentionally left blank...
Page 215: ...10 15 Other Functions 10 This page intentionally left blank...
Page 216: ...10 16 This page intentionally left blank...
Page 217: ...10 17 Other Functions 10 This page intentionally left blank...
Page 218: ...10 18 This page intentionally left blank...
Page 219: ...10 19 Other Functions 10 This page intentionally left blank...