5-2
Explanation
The following table shows the mode, range, and measurable range.
Mode
Range
Measurable Range
Volt
20 mV
–20.00 to 20.00 mV
60 mV
–60.00 to 60.00 mV
200 mV
–200.0 to 200.0 mV
2 V
–2.000 to 2.000 V
6 V
–6.000 to 6.000 V
20 V
–20.00 to 20.00 V
DC current input
An external shunt resistor is connected to the input terminal to convert a current signal to
a voltage signal. The shunt resistors in the following table can be provided. A 250
Ω
shunt resistor, for example, is used to convert a 4 to 20 mA to a 1 to 5 V.
Name
Model (Part) Number
Specification
Shunt resistors
415920
250
Ω
±
0.1%
(for screw terminals)
415921
100
Ω
±
0.1%
415922
10
Ω ±
0.1%
Shunt resistors
438920
250
Ω ±
0.1%
(for clamped terminals)
438921
100
Ω ±
0.1%
438922
10
Ω ±
0.1%