8-8
• Display data and event data
Measurement channel : 2, computation channel : none
Data
Maximum Number of Data Points and Sampling Length
Display data
Maximum number of data points = 900,000/(2 CH
×
4 bytes + 0
×
8 bytes) =
112,500
However, since 75,000 data points is the limit,
Maximum number of data points = 75,000
Display rate When set to 30 min/div (60 s sampling interval)
Sampling length = 75,000 data points
×
60 s = 4,500,000 s (approx. 52 days)
Event data
Maximum number of data points = 300,000/(2 CH
×
2 bytes + 0
×
4 bytes) =
75,000
However, since 30,000 data points is the limit,
Maximum number of data points = 30,000 data points
When the event data sampling interval is 1 s
Sampling length = 30,000 data points
×
1 s = 30,000 s (approx. 8 hours)
Measurement channel : 12, computation channel : 6
Data
Maximum Number of Data Points and Sampling Length
Display data
Maximum number of data points = 900,000/(12 CH
×
4 bytes + 6
×
8 bytes) =
9,375
Display rate When set to 30 min/div (60 s sampling interval)
Sampling length = 9,375 data points
×
60 s = 562,500 s (approx. 6.5 days)
Event data
Maximum number of data points = 300,000/(12 CH
×
2 bytes + 6
×
4 bytes) =
6,250
When the event data sampling interval is 1 s
Sampling length = 6,250 data points
×
1 s = 6,250 s (approx. 1.7 hours)
8.2 Function to Acquire Display Data and Event Data
Summary of Contents for RD-MV100 Series
Page 40: ...1 22 This page intentionally left blank...
Page 41: ...1 23 Overview of Functions 1 This page intentionally left blank...
Page 139: ...7 5 Setting and Operating the Display 7 This page intentionally left blank...
Page 140: ...7 6 This page intentionally left blank...
Page 215: ...10 15 Other Functions 10 This page intentionally left blank...
Page 216: ...10 16 This page intentionally left blank...
Page 217: ...10 17 Other Functions 10 This page intentionally left blank...
Page 218: ...10 18 This page intentionally left blank...
Page 219: ...10 19 Other Functions 10 This page intentionally left blank...