SETUP PROCEDURE
3-87
3.8.10
Confirmation of External Signal (Output)
Check whether the output signal sent to the external equipment connected to the output terminal
can change normally.
Checking is only necessary when output terminal (option) for external signal is installed.
1
Under sensor check mode, turn
the dial to display “EXTERNAL
SIGNAL OUTPUT CHECK”.
2
Press the
key.
Move to setting of BUSY.
3
Turn the dial to display “ON/OFF”.
Setting
Content
ON
Turn ON
OFF
Turn OFF
When switching between ON/OFF, the
external signal changes too. Therefore
please confirm the output signal in the
connected external equipment.
4
Press the
key.
Move to setting of ONLINE.
5
Repeat step 3 and step 4, and
confirm the 4 types of output
signal, namely BUSY, ONLINE,
END and ERROR.
6
After ERROR check, please
press the
key.
Return to selection of setting menu.
Press the
key
to return to selection of
setting mode.