13 - 2
Defibrillator/Monitor Operator’s Manual
13.3
NIBP Measurement Limitations
Measurements are impossible with heart rate extremes of less than 30 bpm or greater than 300 bpm, or if the
patient is on a heart-lung machine.
The measurement may be inaccurate or impossible in the following situations:
■
If a regular arterial pressure pulse is hard to detect;
■
With excessive and continuous patient movement such as shivering or convulsions;
■
With cardiac arrhythmias;
■
Rapid blood pressure changes;
■
Severe shock or hypothermia that reduces blood flow to the peripheries;
■
Obesity, where a thick layer of fat surrounding a limb dampens the oscillations coming from the artery.
13.4
Measurement Modes
There are three modes of measuring NIBP:
■
Manual: measurement on demand.
■
Auto: continually repeated measurements at set intervals.
■
STAT: continually rapid series of measurements over a five minute period, then return to the previous
mode.
13.5
NIBP Display
The NIBP display shows numerics only as below. Your display may be configured to look slightly different.
(1) Measurement mode
(2) NIBP unit: mmHg or kPa
(3) Mean pressure at the completion of measurement, or cuff pressure during the measurement.
(4) Diastolic pressure
(5) NIBP prompt message
(6) Systolic pressure
(7) Time of last measurement
(8) NIBP alarm limits
(8)
(1) (2)
(3)
(6) (5) (4)
(7)
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...