Defibrillator/Monitor Operator’s Manual
21 - 13
21.6.16
Data Management Setup Menu
21.6.17
Test Setup Menu
21.6.18
Network Setup
Alm Rec
HR
On, Off
Off
Sets whether the parameter is
automatically recorded when
the related alarm is triggered.
ARR
PVCs
Resp
SpO2
PR
NIBP
CO2
Menu Item
Options/Range
Default
Description
Menu Item
Options/Range
Default
Description
Tabular Trends Interval
1 min, 2 min, 5 min, 10 min,
15 min, 30 min, 60 min
5 min
Sets the interval that trends are
stored.
Event Wave Length
8s, 16s, 32s
16 s
Sets the length that events are
stored.
Menu Item
Options/Range
Default
Description
User Test Prompt
On, Off
Off
Sets whether the equipment
reminds to perform the user test
if the recommended test time is
reached.
Auto Test Time
24 h: 00 to 23
12 h: 12AM to 11PM
3:00
Sets the initiated time for auto
test performed every day.
Menu Item
Options/Range
Default
Description
Network Type
LAN, WLAN
LAN
Sets the network type.
Address Type
DHCP, Manual
DHCP
•
If [
Address Type
] is set to
[
DHCP
]: the equipment
automatically gets the IP
address.
•
If [
Address Type
] is set
to[
Manual
]: [
Local IP
Address
], [
Subnet Mask
]
and [
Gateway
] are required.
Restoring factory default
configurations does not change
this setting.
Local IP Address
0 - 255
0. 0. 0. 0
Subnet Mask
0 - 255
0. 0. 0. 0
Gateway
0 - 255
0. 0. 0. 0
DNS Address Type
DHCP, Manual
DHCP
•
If [
DNS Address Type
] is set
to [
DHCP
]: the equipment
automatically gets the DNS
address.
•
If [
DNS Address Type
] is set
to[
Manual
]: [
Preferred DNS
Server
] and [
Alternate DNS
Server
] are required.
Preferred DNS Server
0 - 255
0. 0. 0. 0
Alternate DNS Server
0 - 255
0. 0. 0. 0
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...