E - 2
Defibrillator/Monitor Operator’s Manual
E.2
Technical Alarm Messages
ECG
Irr. Rhythm
M*
Arrhythmia has occurred to the patient. Check the patient’s
condition and the ECG connections.
A-Fib
M*
PNP
M*
The pacer appears abnormal. Check the pacer.
PNC
M*
Resp
Resp Apnea
H
The respiration signal was so weak that the equipment cannot
perform respiration analysis. Check the patient’s condition and the
Resp connections.
Resp Artifact
H
The respiration circuit is disturbed. Check for any
possible sources of signal noise.
SpO2
SpO2 Desat
H
The SpO
2
value has fallen below the desaturation alarm limit.
Check the patient’s condition and check if the alarm limit settings
are correct.
No Pulse
L
The pulse signal was so weak that the equipment cannot perform
pulse analysis. Check the patient’s condition, SpO
2
sensor and
measurement site.
CO2
CO2 Apnea
H
The patient stops breathing, or the respiration signal was so weak
that the equipment cannot perform respiration analysis. Check the
patient’s condition, CO
2
accessories and airway connections.
Measurement
Alarm Message
L
Cause and solution
Measurement
Alarm Message
L
I
Cause and solution
XX
XX SelfTest Err
H
C
An error occurred to the XX module, or there is a problem
with the communications between the module and the
main unit
. Restart the equipment.
XX Init Err
H
C
XX Comm Err
L
C
XX Comm Stop
H
C
XX Overrange
L
C
The measured XX value is not within the specified range
for XX measurement. Contact your service personnel.
ECG
ECG Lead Off
L*
B
The ECG electrode has become detached from the patient
or the lead wire has become disconnected from the trunk
cable. Check the connection of the electrodes and
leadwires.
ECG YY Lead Off
(YY represents the
leadwires V, LL, LA, and
RA, as per AHA
standard, or C, F, L and R
as per IEC standard.)
L*
B
Pads/Paddles Off
L*
B
The pads/paddles have been detached from the patient
or the therapy cable is loose. Check that the electrode
pads/paddles and therapy cable are properly connected.
ECG Noise
L
A
The ECG signal is noisy. Check for any possible sources of
signal noise form the area around the cable and electrode,
and check the patient for excessive motion.
ECG Signal Invalid
L
A
ECG amplitude is so low that ECG signal is undetectable.
Check for any possible source of interference from the
area around the cable and electrode; check the patient’s
condition.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...