5 - 12
Defibrillator/Monitor Operator’s Manual
Intermittent Signal
1. Check that cables are properly connected.
2. Check that electrodes are not detached or dry. Perform skin preparation
again as described in
5.4 Preparing for ECG Monitoring and Measurement
.
3. Check that the patient cable or leadwires are not damaged. Change
them if necessary.
Excessive alarms: heart rate, lead fault
1. Check that electrodes are not dry. Perform skin preparation again and
replace the electrodes. For details, refer to
2. Check for excessive patient movement or muscle tremor. Reposition the
electrodes. Replace with fresh and moist electrodes if necessary.
Low Amplitude ECG Signal
1. Check that the ECG gain is not set too low. Adjust the gain as required.
For details, refer to
5.6.5 Changing ECG Waveform Size
.
2. Perform skin preparation again and re-place the electrodes. For more
information, refer to
5.4 Preparing for ECG Monitoring and Measurement
3. Check electrode application sites. Avoid bone or muscular area.
4. Check that electrodes are not dry or used for a prolonged time. Replace
with fresh and moist electrodes if necessary.
No ECG Waveform
1. Check that the ECG gain is not set too low. Adjust the gain as required.
For details, refer to
5.6.5 Changing ECG Waveform Size
.
2. Check that the leadwires and patient cables are properly connected.
Change cable and lead wires.
3. Check that the patient cable or leadwires are not damaged. Change
them if necessary.
Base Line Wander
1. Check for excessive patient movement or muscle tremor. Secure
leadwires and cable.
2. Check that electrodes are not detached or dry and replace with fresh
and moist electrodes if necessary. For details, refer to
ECG Monitoring and Measurement
3. Check for ECG filter setting. Set ECG Filter mode to [
Monitor
].
Problem
Corrective Actions
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...