4 - 6
Defibrillator/Monitor Operator’s Manual
4.6
Pausing Alarms
You can temporarily disable alarm indicators by pressing the Alarm Pause button
on the front panel. When
alarms are paused:
■
For physiological alarms, no alarm indication is shown. New physiological alarm will not be presented.
■
The remaining alarm pause time is displayed in the physiological alarm area.
■
For technical alarms, alarm sounds are paused, but alarm lamps and alarm messages remain presented.
■
The alarm pause symbol
is displayed in the sound symbol area. If a new technical alarm is triggered in
the alarm paused period, the alarm message will be displayed.
When the alarm pause time expires, the alarm paused status is automatically deactivated. You can also cancel
the alarm paused status by pressing the Alarm Pause button
.
The default alarm pause time is 2 minutes. To change alarm pause time:
1.
Press the Main Menu button on the front panel, and then select [
Others >>
]
→
[
Configuration >>
]
→
enter the required password.
2.
Select [
Alarm Setup
]
→
[
Alarm Pause Time
] and then select an appropriate value.
4.7
Switching Alarms Off
When an alarm is switched off, the alarm status is the same with that when an alarm is paused.
Alarms are switched off when:
■
The Alarm Pause button
is pressed if the [
Alarm Pause Time
] is set to [
Permanent
].
■
The operating mode is switched to the Manual Defib mode.
■
Synchronized cardioversion is exited when operating in the Manual Defib mode.
The alarm off status is exited when:
■
The Alarm Pause button
is pressed.
■
Sync Defib is switched on in the Manual Defib mode.
4.8
Switching Off Alarm Sounds
In the event that [
Audio Off
] is enabled through the Configuration Main menu, to switch off the alarm tone, set
[
Alm Volume
] to 0 while operating in the Monitor mode, Manual Defib mode or Pacer mode. In the audio off
status, the audio off symbol
appears in the sound symbol area. In this case, the alarm status is the same with
that when alarm tones are paused.
The audio off status is exited when:
■
[
Audio Off
] is set to [
Disabled
] by selecting [
Alarm Setup
] from the Configuration Management Main
menu.
■
The Alarm Pause button
is pressed. In this case, the equipment enters the alarm paused status and
the alarm volume is reset to the default level. the alarm pause symbol
is displayed in the sound
symbol area.
■
Operating mode is switched.
■
Alarm volume is changed to a value between 1 and 10.
4.9
Resetting Alarms
By pressing the [
Alarm Reset
] soft key, you can reset the alarm system and enable the alarm system to respond
to a subsequent alarm condition.
For physiological alarms, when the alarm system is reset:
■
The alarm sound is reset.
■
A √ appears before the alarm message.
■
The alarm reset symbol
appears in the alarm symbol area.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...