A - 8
Defibrillator/Monitor Operator’s Manual
Pace Pulse
Pace pulse markers
Pace pulses meeting the following conditions are labelled with a PACE
marker:
Amplitude:
Width:
Rise time:
±2 to ± 700 mV
0.1 to 2 ms
10 to 100 μs
Pace pulse rejection
When tested in accordance with the IEC 60601-2-27: 201.12.1.101.13, the
heart rate meter rejects all pulses meeting the following conditions.
Amplitude:
Width:
Rise time:
±2 to ± 700 mV
0.1 to 2 ms
10 to 100 μs
HR
Measurement range
Pediatric:
Adult:
15 to 350 bpm
15 to 300 bpm
Accuracy
±1% or ±1bpm, which ever is greater
Resolution
1 bpm
Sensitivity
200 μV
Heart rate averaging
In compliance with the requirements in Clause 201.7.9.2.9.101 b) 3) of IEC
60601-2-27, the following method is used:
If the last 3 consecutive RR intervals are greater than 1200 ms, the 4 most
recent RR intervals are averaged to compute the HR. Otherwise, heart rate is
computed by subtracting the maximum and minimum ones from the most
recent 12 RR intervals and then averaging them.
The HR value displayed on the screen is updated every second.
Response time to heart rate change
Meets the requirements of IEC 60601-2-27: Clause 201.7.9.2.9.101 b) 5).
From 80 to 120 bpm:
From 80 to 40 bpm:
less than 11 s
less than 11 s
Time to alarm for tachycardia
Meets the requirements in Clause 201.7.9.2.9.101 b) 6) of IEC 60601-2-27.
Waveform
4ah - range:
4a - range:
4ad - range:
4bh - range:
4b - range:
4bd - range:
<11 s
<11 s
<11 s
<11 s
<11 s
<11 s
Arrhythmia Analysis Classifications
Asystole, V-Fib/V-Tach, PNP, PNC
Tall T-wave rejection capability
When the test is performed based on Clause 201.12.1.101.17 of IEC 60601-2-
27, the heart rate meter will reject all 100 ms QRS complexes with less than
1.2 mV of amplitude, and T waves with T-wave interval of 180 ms and those
with Q-T interval of 350 ms.
Lead-off detection current
≤0.1 μA
Baseline recovery time
<2.5 s (after defibrillation, in therapy mode)
Response to irregular rhythm
In compliance with the requirements in Clause 201.7.9.2.9.101 b) 4) of IEC
60601-2-27, the heart rate after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy (3a): 80±1 bpm
Slow alternating ventricular bigeminy (3b): 60±1 bpm
Rapid alternating ventricular bigeminy (3c): 120±1 bpm
Bidirectional systoles (3d): 90±2 bpm
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...