Defibrillator/Monitor Operator’s Manual
6 - 5
WARNING
•
The CPR metronome sounds do not indicate information regarding the patient’s condition. Because
patient status can change in a short time, the patient should be assessed at all times. Do not perform
CPR on a patient who is responsive or is breathing normally.
NOTE
•
The CPR metronome is deactivated when [Voice Prompts] in the
[
AED Setup
]
menu is configured off
through the Configuration Main menu.
•
The volume of CPR metronome is affected by [Voice Volume] in the
[
AED Setup
]
menu configured
through the Configuration Main menu.
6.8
AED Sound Recording
The equipment provides a sound recording function that can record the voice information during AED therapy.
The sound recording function can be configured on or off.
To switch on or off the sounding recording:
1.
Switch the Mode Select knob to Monitor, Manual Defib or Pacer.
2.
Press the Main Menu button on the front panel, and then select
→
[
Others >>
]
→
[
Configuration >>
]
→
enter the required password.
3.
Select [
AED Setup
]
→
[
Voice Recording
], and toggle between [
On
] and [
Off
].
is shown at the top right corner of the AED information area if the sounding recording function is
enabled.
The equipment can store up to 180 minutes, maximum 60 minutes for one patient, of sound recording.
6.9
AED Setup
1.
Switch the Mode Select knob to Monitor, Manual Defib or Pacer.
2.
Press the Main Menu button on the front panel, and then select
→
[
Others >>
]
→
[
Configuration >>
]
→
e
nter the required password.
3.
Select [
AED Setup
] , and then change AED settings as desired.
For details, refer to
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...