3 - 4
Defibrillator/Monitor Operator’s Manual
4.
Rotate the Navigation knob to move the cursor on [
Ok
], and then press the Navigation knob to save the
information inputted.
■
Changing settings
Changing the ECG lead in the Monitor mode is taken as an example below:
1.
Rotate the Navigation knob to move the cursor on the ECG lead label
2.
Press the Navigation knob to highlight the selection
.
3.
Rotate the Navigation knob until you find the desired item, and then press the Navigation knob to confirm
the selection.
3.2.7
Editing Current Patient Information
The patient name/bed number (configurable), patient category and paced status of the current patient are
displayed in the upper left corner of the main screen.
1.
Press the Main Menu button on the front panel, and then select [
Patient Demographics >>
].
2.
Make changes as desired.
◆
Set [
Patient Cat.
] to change the patient category. You can also rotate the Navigation knob to move
the cursor on the patient category of the main screen, press and rotate the Navigation knob to
change the patient category.
◆
Edit patient name or bed number. You can set the patient information type displayed in the upper left
corner of the main screen by selecting [
General Setup
] through the Configuration Main menu.
◆
Set paced status: when the [
Paced
] is set to [
Yes
], a paced symbol
is displayed in the upper left
corner of the main screen.
3.
Select [
Others >>
] to edit more patient information.
If the equipment is connected to the CMS, the current patient name, bed number, MRN, department, height,
weight, admit date or doctor information can also be changed on the CMS.
3.2.8
Turning Off the Equipment
1.
Check that the patient monitoring or therapy is completed.
2.
Disconnect the patient cables and sensors from the patient.
3.
Make sure to save or clear the patient data as required.
4.
Turn the Mode Select Knob to Off. After 10 seconds, the equipment is shut down.
NOTE
•
To completely disconnect the power supply, unplug the power cord.
3.2.9
Auto Restoring to Last Configuration
During operation, you may make changes to some settings. However, these changes may not be saved as user
configuration. To prevent the changes from losing in case of sudden power failure, the equipment saves the
settings in real time. The saved settings are the latest configuration. In case of power failure, the equipment
loads the latest configuration if restarts within 60 seconds; it load the user configuration if restarts 120 seconds
later after the power failure; it may load either the latest configuration or the user configuration if restarts
between 60 and 120 seconds after the power failure.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...