Defibrillator/Monitor Operator’s Manual
5 - 9
*: indicates that this arrhythmia alarm is not presented when [
Paced
] is set to [
Yes
].
**: indicates that this arrhythmia alarm is not presented when [
Paced
] is set to [
No
].
■
When electrode pads are used for ECG monitoring, the equipment provides only 4 arrhythmia alarms,
including asystole, ventricular fibrillation/ventricular tachycardia, PNP, and PNC.
■
When external paddles are used, the equipment provides only 3 arrhythmia alarms, including ventricular
fibrillation/ventricular tachycardia, PNP, and PNC.
5.7.2
Switching Arrhythmia Analysis On and Off
To switch arrhythmia analysis on or off:
1.
Select the ECG parameter area to enter the [
ECG Setup
] menu,.
2.
Select [
Arrhythmia >>
]
→
[
Arrhythmia
], and toggle between [
On
] and [
Off
].
You can also swtich on or off [
Arrhythmia
] by selecting [
ECG Setup
] from the Configuration Main menu.
5.7.3
Changing Arrhythmia Alarm Settings
To change arrhythmia alarm settings:
1.
Select the ECG parameter area to enter the [
ECG Setup
] menu
2.
Select [
Arrhythmia >>
]
→
[
Arrh. Alarm >>
], where you can set alarm switch, alarm level and alarm record
switch for all the arrhythmia events.
You can also set [
ARR Alm Lev
] by selecting [
ECG Setup
] from the Configuration Main menu.
PVCs/min
PVCs/min exceeds high limit
Nonlethal
arrhythmia
PNP**
No pace pulse detected for 1.75 x average R-to-R intervals following a
QRS complex (for paced patients only).
PNC**
No QRS complex detected for 300 milliseconds following a pace pulse (for
paced patients only).
PVC
One PVC detected in normal heartbeats
Couplet
Paired PVCs are detected.
VT>2
More than 2 consecutive PVCs within the last minute.
Bigeminy
A dominant rhythm of N, V,N, V, N, V.
Trigeminy
A dominant rhythm of N, N, V,N, N, V, N, N, V.
R ON T
R on T detected in normal heartbeats.
Missed Beat*
No beat detected for 1.75x average R-R interval for HR <120, or No beat
for 1 second with HR >120 (for non-paced patients only), or No beat
detected for more than the set pause threshold.
Brady
The average heart rate is less than 60 bpm.
Tachy
The average heart rate is greater than 100 bpm.
Vent Rhythm
The consecutive PVCs > the Vbrd PVCs limit, and the HR is between Vbrd
Rate limit and the Vtac Rate limit.
Multif. PVCs
Multiform PVCs detected in Multif. PVC's Window (which is adjustable).
Nonsus. Vtac
The consecutive PVCs < the Vtac PVCs limit but > 2, and HR > the Vtac
Rate limit.
Pause*
No QRS detected within the set time threshold of pause.
Irr. Rhythm
Consistently irregular rhythm.
A-Fib (for adult only)
P wave is absent and normal beat RR intervals are irregular.
Arrhythmia event
Description
Category
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...