Defibrillator/Monitor Operator’s Manual
7 - 1
7
Manual Defibrillation
7.1
Manual Defibrillation Introduction
This chapter explains how to prepare for and perform asynchronous defibrillation and synchronized
cardioversion using electrode pads, external paddles and internal paddles.
In the Manual Defib Mode, you must assess the ECG waveforms, decide if defibrillation or cardioversion is
indicated, select appropriate energy setting, charge the equipment, and deliver the shock. Text messages and a
contact impedance indicator on the screen provide relevant information to guide your throughout the
defibrillation process.
While operating manual defibrillation, besides ECG, you can select to monitor parameters from SpO
2
, NIBP and
CO
2
. Alarms are turned off automatically when you enter the Manual Defib mode. Pressing the Alarm Pause
button
on the front panel of the equipment can turn on the alarms.
7.2
Manual Defibrillation Safety Information
DANGER
•
Defibrillation current can cause operator or bystander severe injury or even death. Never touch the
patient or any metal objects connected to the patient (including the bed or gurney) during
defibrillation.
•
Avoid contact between parts of the patient’s body such as exposed skin of head or limbs, conductive
fluids such as gel, blood, or saline, and metal objects such as a bed frame or a stretcher which may
provide unwanted pathways for the defibrillating current.
•
Do not allow electrode pads and paddles to touch each other or to touch other ECG monitoring
electrodes, lead wires, dressings, etc. Contact with metal objects may cause electrical arcing and
patient skin burns during defibrillation and may divert current away from the heart.
•
During manual defibrillation, make sure your hands are dry and free from conductive gel to avoid
shock hazard.
•
Use care when operating this equipment close to oxygen sources (such as bag-valve-mask devices or
ventilator tubing). Turn off gas source or move source away from patient during defibrillation. This
can cause an explosion hazard.
WARNING
•
During synchronized cardioversion, if monitoring patient’s ECG through external paddles, artifact
introduced by paddle movement may resemble an R-wave and trigger a defibrillation shock.
•
Do not use conductive liquid. Use only conductive gel specified by the equipment manufacturer.
•
If external paddles are used for defibrillation, apply the external paddles tightly and evenly to the
patient’s chest to ensure good skin contact.
•
Never apply the paddles to human body to verify paddle connection.
•
Clinicians must select an appropriate energy level for defibrillation of pediatric patients.
•
Never charge and deliver shock frequently in non-clinical situations. Otherwise equipment damage
could occur.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...