Defibrillator/Monitor Operator’s Manual
5 - 7
■
The P-waves and T-waves should be less than 0.2mV.
5.6.4
Setting the ECG Waveform Layout
1.
Select the ECG parameter area to enter the [
ECG Setup
] menu.
2.
Select [
Cascade
] and toggle between [
On
] and [
Off
].
5.6.5
Changing ECG Waveform Size
If the ECG waveform is too small or clipped, you change its size by using the Navigation knob.
1.
Rotate the Navigation knob to move the cursor on the ECG waveform gain.
2.
Press the Navigation knob to highlight it.
3.
Rotate the Navigation knob until you find the desired item, and then press the knob to confirm the
selection.
5.6.6
Changing ECG Waveform Speed
1.
Select the ECG parameter area to enter the [
ECG Setup
] menu.
2.
Select [
Sweep
], and then choose an appropriate value. The faster the wave sweeps, the wider the wave is.
You can also set [
Sweep
] by selecting [
ECG Setup
] from the Configuration Main menu.
5.6.7
Setting the ECG Filter
■
When monitoring ECG through 3/5- lead set, filter mode is displayed above the first ECG waveform. The
available filter mode settings are [
Monitor
], [
Therapy
], and [
Diagnostic
].
■
When monitoring ECG through external paddles/electrode pads, filter mode is always [
Therapy
] and can
not be changed.
To set the ECG filter mode:
1.
Rotate the Navigation knob to move the cursor on the ECG filter mode.
2.
Press the Navigation knob to highlight it.
3.
Rotate the Navigation knob until you find the desired item, and then press the knob to confirm the
selection.
NOTE
•
For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the system may
mistake an internal pace pulse for a QRS or fail to alarm when the pacer is broken.
5.6.8
Switching On or Off the Notch Filter
The notch filter removes the AC line noise. Switching on the notch filter is recommended when there is
interference with the waveform.
■
When the ECG filter mode is set to [
Monitor
] or [
Therapy
], [
Notch Filter
] is always [
On
].
■
When the ECG filter mode is set to [
Diagnostic
], [
Notch Filter
] can be switched on or off as required.
Set notch frequency according to the electric power frequency of your country. To switch on or off notch filter:
1.
Select the ECG parameter area to enter the [
ECG Setup
] menu.
2.
Select [
Others >>
]
→
[
Notch Filter
], and toggle between [
On
] and [
Off
].
You can also set the power frequency of notch filter by selecting [
ECG Setup
] through the Configuration Main
menu.
NOTE
N
•
The setting of
[
Notch Filter
]
will not be changed by restoring to factory default settings nor shutting
down the system.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...