Defibrillator/Monitor Operator’s Manual
A - 9
A.5.3
Resp Specifications
A.5.4
SpO
2
Specifications
Mindray SpO
2
Module
Masimo SpO
2
Module
Technique
Trans-thoracic impedance
Measurement range
0 to 200 rpm
Resolution
1 rpm
Accuracy
121 to 200 rpm:
0 to 120 rpm:
±2 rpm
±1 rpm
Respiration excitation waveform
<300 μA, sinusoid, 62.8 kHz (±10%)
Minimum respiration impedance
threshold
0.3Ω
with× 5 gain
Bandwidth
0.2 to 2.5 Hz (-3 dB)
Reference impedance range
2200 to 4500Ω, using an ECG cable with 1 kΩ resistor
Apnea alarm time
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Standard
Meet standards of ISO 80601-2-61
Measurement range
0 to 100%
Resolution
1%
Response time
<20 s (SpO
2
value sudden changes from 70% to 100%)
Accuracy
70 to 100%:
70 to 100%:
0% to 69%:
±2% (in adult/pediatric mode)
±3% (in neonate mode)
Not specified
Refreshing rate
≤2 s
*Measurement accuracy verification: The SpO
2
accuracy has been verified in human experiments by comparing with
arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed
and about two-thirds of the measurements are expected to come within the specified accuracy range compared to CO-
oximeter measurements.
Standard
Meet standards of ISO 80601-2-61
Measurement range
1 to 100%
Resolution
1%
Response time
≤20 s (SpO
2
value sudden changes from 70% to 100%)
Accuracy
1
70 to 100%:
±2% (measured without motion in adult/
pediatric mode)
70 to 100%:
±3% (measured without motion in neonate
mode)
70 to 100%:
±3% (measured with motion)
1% to 69%:
Not specified
Refreshing rate
≤2 s
SpO
2
averaging time
2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...