Defibrillator/Monitor Operator’s Manual
4 - 5
NOTE
•
You can enable auto alarm limits only when the current parameter measurement is within the auto
alarm limits range.
Module
Parameter
Low alarm limit
High alarm limit
Auto alarm limits range
Adult/pediatric
Neonate
Adult/pediatric
Neonate
ECG
HR
HR×0.8 or 40
bpm (whichever
is greater)
(HR-30) or 90
bpm (whichever
is greater)
HR×1.25 or 240
bpm whichever
is lower)
(HR + 40) or 200
bpm whichever
is lower)
Adult/pediatric: 35 to 240
Neonate: 55 to 225
Resp
RR
RR×0.5 or 6/min
(whichever is
greater)
(RR-10) or 30/
min (whichever
is greater)
RR×1.5 or 30/
min (whichever
is lower)
(RR+25) or 85/
min (whichever
is lower)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
SpO
2
SpO
2
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
PR
PR×0.8 or 40
bpm (whichever
is greater)
(PR-30) or 90
bpm (whichever
is greater)
PR×1.25 or 240
bpm whichever
is lower)
(PR + 40) or 200
bpm whichever
is lower)
Adult/pediatric: 35 to 240
Neonate: 55 to 225
NIBP
NIBP-S
SYS×0.68 +
10mmHg
(SYS-15) or
45mmHg
(whichever is
greater)
SYS×0.86 +
38mmHg
(SYS + 15) or
105mmHg
(whichever is
lower)
Adult: 45 to 270
Pediatric: 45 to 185
Neonate: 35 to 115
NIBP-D
Dia×0.68 +
6mmHg
(Dia-15) or
20mmHg
(whichever is
greater)
Dia×0.86 +
32mmHg
(Dia + 15) or
80mmHg
(whichever is
lower)
Adult: 25 to 225
Pediatric: 25 to 150
Neonate: 20 to 90
NIBP-M
Mean×0.68 +
8mmHg
(Mean-15) or
35mmHg
(whichever is
greater)
Mean×0.86 +
35mmHg
(Mean + 15) or
95mmHg
(whichever is
lower)
Adult: 30 to 245
Pediatric: 30 to 180
Neonate: 25 to 105
CO
2
EtCO
2
0-32mmHg:
remains
unchanged
0-32mmHg:
remains
unchanged
0-32mmHg:
remains
unchanged
0-32mmHg:
remains
unchanged
Same as the
measurement range
32-35mmHg:
29mmHg
32-35mmHg:
29mmHg
32-35mmHg:
41mmHg
32-35mmHg:
41mmHg
35-45mmHg:
EtCO
2
-6mmHg
35-45mmHg:
EtCO
2
-6mmHg
35-45mmHg:
EtCO
2
+6mmHg
35-45mmHg:
EtCO
2
+6mmHg
45-48mmHg:
39mmHg
45-48mmHg:
39mmHg
45-48mmHg:
51mmHg
45-48mmHg:
51mmHg
>48mmHg:
remains
unchanged
>48mmHg:
remains
unchanged
>48mmHg:
remains
unchanged
>48mmHg:
remains
unchanged
FiCO
2
N/A
N/A
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
awRR
awRR×0.5 or 6/
min (whichever
is greater)
awRR-10 or 30/
min (whichever
is greater)
awRR×1.5 or 30/
min (whichever
is lower)
awRR+25 or 85/
min (whichever
is lower)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...