A - 6
Defibrillator/Monitor Operator’s Manual
Common mode rejection
Diagnostic mode:
Monitor mode:
Therapy mode:
>90 dB
>105 dB
>105 dB
Notch filter
50/60Hz,
In Monitor
and Therapy modes: notch filter turns on automatically
In
Diagnostic mode: notch filter is turned on manually
ECG signal range
±8mV (peak-to-peak value)
Calibration signal
1mV (peak-to-peak value) ±5%
Differential input impedance
≥5 MΩ
Electrode offset potential tolerance
±500mV
Defibrillation protection
Enduring 5000V (360 J) charge without data loss or corruption
Baseline recovery time: <2.5 s (after defibrillation)
Polarization recovery time: <10 s
Defibrillation energy absorption: ≤10% (100Ω load)
ESU protection
Cut mode: 300 W
Coagulate mode: 100 W
Recovery time: ≤10 s
In compliance with the requirements in clause 202.6.2.101 of
IEC 60601-2-27
Pace Pulse
Pace pulse markers
Pace pulses meeting the following conditions are labelled with a PACE
marker:
Amplitude:
Width:
Rise time:
±2 to ± 700 mV
0.1 to 2 ms
10 to 100 μs
Pace pulse rejection
When tested in accordance with the IEC 60601-2-27: 201.12.1.101.13, the
heart rate meter rejects all pulses meeting the following conditions.
Amplitude:
Width:
Rise time:
Input slew rate:
±2 to ± 700 mV
0.1 to 2 ms
10 to 100 μs
2.2 V/s ± 15% RTI
HR
Measurement range
Neonate:
Pediatric:
Adult:
15 to 350 bpm
15 to 350 bpm
15 to 300 bpm
Accuracy
±1% or ±1bpm, which ever is greater
Resolution
1 bpm
Sensitivity
200 μV (lead II)
Heart rate averaging
In compliance with the requirements in Clause 201.7.9.2.9.101 b) 3) of IEC
60601-2-27, the following method is used:
If the last 3 consecutive RR intervals are greater than 1200 ms, the 4 most
recent RR intervals are averaged to compute the HR. Otherwise, heart rate is
computed by subtracting the maximum and minimum ones from the most
recent 12 RR intervals and then averaging them.
The HR value displayed on the screen is updated every second.
Response time to heart rate change
Meets the requirements of IEC 60601-2-27: Clause 201.7.9.2.9.101 b) 5).
From 80 to 120 bpm:
From 80 to 40 bpm:
less than 11 s
less than 11 s
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...