Defibrillator/Monitor Operator’s Manual
4 - 3
■
For technical alarms
4.3.4
Flashing Numerics
If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in alarm will flash every
second, and corresponding alarm limit will also flash at the same frequency indicating the alarm limit is violated.
4.3.5
Alarm Status Symbols
Apart from the aforementioned alarm indicators, the equipment still uses the following symbols telling the
alarm status:
4.4
Alarm Tone Configuration
4.4.1
Changing the Alarm Volume
1.
Press the Main Menu button on the front panel, and then select [
Alarm Setup >>
] .
2.
Set [
Alm Volume
] to an appropriate level.
◆
If [
Audio Off
] is enabled, alarm volume can be set to a value between 0 and 10, in which 0 means
audio off and 10 the maximum volume level.
◆
If [
Audio Off
] is disabled, alarm volume can be set to a value between 1 and 10, in which 1 is the
minimum volume level and 10 the maximum.
The setting of alarm volume will not be saved when the equipment is powered off.
You can also disable or enable [
Audio Off
], set the alarm volume by selecting [
Alarm Setup
] from the
Configuration Main menu. In this case, the setting will be saved.
NOTE
E
•
You cannot adjust the alarm volume when an alarm is switched off.
4.4.2
Setting the Reminder Tones
When alarms or alarm sounds are turned off, the equipment gives a reminder tone of a single beep every 60
seconds.
The reminder tone is switched off by default. You can switch on [
Reminder Tone
] by selecting [
Alarm Setup
]
from the Configuration Main menu. You can also change the reminder volume. The default reminder volume is
[
Med
].
4.4.3
Setting the Interval between Alarm Sounds
1.
Press the Main Menu button on the front panel, and then select [
Others >>
]
→
[
Configuration >>
]
→
enter the required password.
2.
Select [
Alarm Setup
].
3.
Respectively set [
High Level Alarm (s)
], [
Med Level Alarm (s)
] and [
Low Level Alarm (s)
].
◆
High level alarms:
red
◆
Medium level alarms:
yellow
◆
Low level alarms:
blue
Alarm pause:
indicates alarms are paused.
Alarm reset:
indicates alarm are reset.
Audio off:
indicates alarm sounds are turned off.
Alarm off:
indicates the system is in alarm off status.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...