Defibrillator/Monitor Operator’s Manual
3 - 1
3
Basic Operations and Settings
3.1
Equipment Installation
WARNING
•
The equipment shall be installed by personnel authorized by the manufacturer.
•
The software copyright of the equipment is solely owned by the manufacturer. No organization or
individual shall resort to juggling, copying, or exchanging it or to any other infringement on it in any
form or by any means without due permission.
•
Devices connected to the equipment must meet the requirements of the applicable IEC standards
(e.g. IEC 60950 safety standards for information technology equipment and IEC 60601-1 safety
standards for medical electrical equipment). The system configuration must meet the requirements
of the IEC 60601-1 medical electrical systems standard. Any personnel who connect devices to the
equipment’s signal input/output port is responsible for providing evidence that the safety
certification of the devices has been performed in accordance to the IEC 60601-1. If you have any
questions, please contact the manufacturer.
•
If it is not evident from the equipment specifications whether a particular combination is hazardous,
for example, due to summation of leakage currents, consult the manufacturers or else an expert in
the field, to ensure the necessary safety of all devices concerned will not be impaired by the
proposed combination.
CAUTION
•
The docking station is part of the equipment. Use only the specified docking station.
3.1.1
Unpacking and Checking
Before unpacking, examine the packing case carefully for signs of damage. If any damage is detected, contact
the carrier or the manufacturer. If the packing case is intact, open the package and remove the equipment and
accessories carefully. Check all materials against the packing list and check for any mechanical damage. If you
have any question, please contact us.
WARNING
•
When disposing of the packaging material, be sure to observe the applicable waste control
regulations and keep it out of children’s reach.
•
The equipment might be contaminated during storage and transport. Before use, please verify
whether the packages are intact, especially the packages of single use accessories. In case of any
damage, do not apply it to patients.
NOTE
•
Save the packing case and packaging material as they can be used if the equipment must be
reshipped.
3.1.2
Environmental Requirements
The operating environment of the equipment must meet the requirements specified in this manual.
The environment where the equipment is used shall be reasonably free from noises, vibration, dust, corrosive,
flammable and explosive substances. If the equipment is installed in a cabinet, sufficient space in front and
behind shall be left for convenient operation, maintenance and repair. Moreover, to maintain good ventilation,
the equipment shall be at least 2 inches (5 cm) away from around the cabinet.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...