12 - 2
Defibrillator/Monitor Operator’s Manual
CAUTION
•
Change the application site or replace the sensor and/or patient cable when a persistent SpO2 Low
Signal Quality message is displayed on the equipment. These messages may indicate that patient
monitoring time is exhausted on the patient cable or sensor.
•
Replace the cable or sensor when a “SpO2 Sensor Off”, “SpO2 No Sensor”, or “SpO2 Low Signal
Quality” message is consistently displayed while monitoring consecutive patients after completing
troubleshooting steps listed in this manual.
•
Variation in measurements may be profound and may be affected by sampling technique as well as
the patient's physiological conditions. Any results exhibiting inconsistency with the patient’s clinical
status should be repeated and/or supplemented with additional test data. Blood samples should be
analyzed by laboratory instruments prior to clinical decision making to completely understand the
patient’s condition.
•
Use only SpO
2
sensors specified in this manual. Follow the SpO
2
sensor’s instructions for use and
adhere to all warnings and cautions.
•
Do not place the pulse oximeter where the controls can be changed by the patient.
•
If using pulse oximetry during full body irradiation, keep the sensor out of the radiation field. If the
sensor is exposed to the radiation, the reading might be inaccurate or the device might read zero for
the duration of the active irradiation period.
NOTE
•
Additional information specific to the Masimo sensors compatible with the equipment, including
information about parameter/measurement performance during motion and low perfusion, may be
found in the sensor's directions for use (DFU).
•
Masimo cables and sensors are provided with X-Cal™ technology to minimize the risk of inaccurate
readings and unanticipated loss of patient monitoring. Refer to the Cable or Sensor DFU for the
specified duration of the patient monitoring time.
•
The SpO
2
extension cable should be compatible with the SpO
2
connectors. For example, you can
only connect the Mindray SpO
2
extension cable to the Mindray SpO
2
connectors.
•
The SpO
2
simulator can be used to verify the accuracy of the SpO
2
sensor.
12.4
SpO
2
Measurement Limitations
The following factors may influence the accuracy of SpO
2
measurement:
■
Patient physiological characteristics:
◆
Cardiac arrest
◆
Hypotension
◆
Darkly pigmented skin
◆
Shock
◆
Severe vasoconstriction
◆
Hypothermia
◆
Severe anemia
◆
Ventricular septal defects (VSDs)
◆
Venous pulsations
◆
Poor perfusion
◆
Dysfunctional hemoglobin, such as carboxyhemoglobin (COHb) and methemoglobin (MetHb)
◆
Elevated levels of bilirubin
◆
Vasospastic disease, such as Raynaud’s, and peripheral vascular disease
◆
Hemoglobinopathies and synthesis disorders such as thalassemias, Hb s, Hb c, sickle cell, etc.
◆
Hypocapnic or hypercapnic conditions
◆
Birthmark(s), tattoos, skin discolorations, moisture on skin, deformed or abnormal fingers. etc.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...