Defibrillator/Monitor Operator’s Manual
A - 13
Note: The specifications above are based on a new battery, and at 20ºC±5 °C of ambient temperature.
Battery LI24I001A charge
time
Charged by the equipment
connected to the AC power
Less than 2 hours to 90% and less than 3 hours to 100%
with equipment power off;
Less than 3.5 hours to 90% and less than 4.5 hours to
100% with equipment power on.
Charged by the charger station
Less than 2.5 hours to 90% and less than 3 hours to
100%.
Operating
mode
One new fully
charged
battery
Testing condition
Battery LI24I005A run time
Monitoring
≥6 h
The equipment is configured with a 5-lead ECG, Resp,
SpO
2
and CO
2
, NIBP measurements set at an interval of
15 minutes. WI-Fi is disabled. The screen brightness is
set to the factory default without recording.
Defibrillation
≥200
discharges
360J discharge, the equipment is configured with a 5-
lead ECG, Resp, SpO
2
, NIBP and CO
2
. WI-Fi is disabled.
The screen brightness is set to the factory default
without recording.
Defibrillation
≥300
discharges
200J discharge, the equipment is configured with a 5-
lead ECG, Resp, SpO
2
, NIBP and CO
2
. WI-Fi is disabled.
The screen brightness is set to the factory default
without recording.
Pacing
≥4.5 h
50 Ω load impedance, pacing rate 80ppm, pacing
output 60mA. The equipment is configured with a 5-
lead ECG, Resp, SpO
2
, NIBP and CO
2
. WI-Fi is disabled.
The screen brightness is set to the factory default
without recording.
Operating
mode
One new fully
charged
battery
Testing condition
Battery LI24I001A run time
Monitoring
≥3 h
The equipment is configured with a 5-lead ECG, Resp,
SpO
2
and CO
2
, NIBP measurements set at an interval of
15 minutes.
WI-Fi is disabled. The screen brightness is
set to the factory default without recording.
Defibrillation
≥100
discharges
360J discharge at intervals of one minute, without
recording
Pacing
≥2 h
50 Ω load impedance, pacing rate 80ppm, pacing
output 60mA, without recording
Battery fuel gauge
5 LEDs indicating the current battery charge level
Shutdown delay
At least 20 minutes of monitoring and six 360J discharges (after the low battery alarm
occurs)
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...