Defibrillator/Monitor Operator’s Manual
10 - 1
10
Monitoring Resp
10.1
Resp Introduction
Impedance respiration is measured across the thorax. When the patient is breathing or ventilated, the volume of
air changes in the lungs, resulting in impedance changes between the electrodes. Respiration rate (RR) is
calculated from these impedance changes, and a respiration waveform appears on the equipment screen.
Resp monitoring is intended for adult, pediatric and neonatal patients.
10.2
Resp Safety Information
WARNING
•
When monitoring the patient’s respiration, do not use ESU-proof ECG cables.
•
The respiration measurement does not recognize obstructive and mixed apneas: it only indicates an
alarm when a pre-adjusted time had elapsed since the last detected breath. The safety and
effectiveness of the respiration measurement method in the detection of apnea, especially the
apnea of prematurity and apnea of infancy, has not been established.
10.3
Resp Display
10.4
Placing Resp Electrodes
As the skin is a poor conductor of electricity, preparing the skin is necessary for a good respiration signal. For
details, refer to
5.4.1 Preparing the Patient Skin
As the Respiration measurement adopts the standard ECG electrode placement, you can use different ECG
cables (3-lead or 5-lead). Since the respiration signal is measured between two ECG electrodes, if a standard ECG
electrode placement is applied, the two electrodes should be RA and LA of ECG Lead I, or RA and LL of ECG Lead
II.
(1) Resp waveform gain
(2) Resp lead label
(3) Resp alarm limits
(4) Respiration rate (RR)
(5) RR unit
(1) (2)
(3)
(5)
(4)
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...