13 - 4
Defibrillator/Monitor Operator’s Manual
WARNING
•
Continuous non-invasive blood pressure measurements may cause purpura, ischemia and
neuropathy in the limb with the cuff. Inspect the application site regularly to ensure skin quality and
inspect the extremity of the cuffed limb for normal color, warmth and sensitivity. If any abnormity
occurs, move the cuff to another site or stop the blood pressure measurements immediately.
13.6.6
Starting a STAT Measurement
1.
Select the NIBP parameter area to enter the [
NIBP Setup
] menu.
2.
Select [
NIBP STAT
]. The STAT mode initiates a 5-minute continuous, automatic NIBP measurement.
13.7
Setting Initial Cuff Inflation Pressure
You can set the initial cuff inflation pressure manually.
1.
Select the NIBP parameter area to enter the [
NIBP Setup
] menu.
2.
Select [
Initial Pressure
] and then select the appropriate setting. The cuff is inflated accordingly at the next
NIBP measurement.
NOTE
•
For known hypertensive patients, you need to set initial cuff pressure to a higher value to reduce the
measurement time.
•
Setting initial cuff inflation pressure is disabled during NIBP measurement.
•
The initial cuff inflation pressure is restored to the default setting if NIBP module has been reset or
patient category has been changed.
13.8
Setting Pressure Unit
You can set [
Press. Unit
] to change NIBP unit only by selecting [
NIBP Setup
] from the Configuration Main menu.
13.9
NIBP Troubleshooting
.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...