Defibrillator/Monitor Operator’s Manual
24 - 3
24.3.2
User Test
If the auto test fails, the user test must be performed to clear the faults.
When turned on, the equipment automatically checks the time to Routine Test, Energy Delivery Test and
Controls Test performed last time. The equipment can be configured to give a “
User Test Due
” message to
remind performing the user test. You can enable it by selecting [
Test Setup
] through the Configuration Main
menu.
WARNING
•
Do not perform the user test when a patient is connected to the equipment.
NOTE
•
Before the user test or after each use, thoroughly clean the external paddles and properly place
them in the paddle tray. The user test passes only when external paddles properly contact the metal
parts of the paddle tray.
•
If the impedance value indicated by impedance indicator changes greatly, check that external
paddles and metal parts of the paddle tray are clean.
•
Install at least one battery and properly place the external paddles in the paddle tray or connect the
pads cable and 50 Ω test load. Otherwise the user test will fail.
24.3.2.1
Starting the User Test
You can start the user test mode while operating in the Monitor, Manual Defib or Pacer mode. Patient
monitoring and therapy automatically end when you enter the User Test Main menu.
1.
Press the Main Menu button on the front panel, and then select [
User Test >>
]
→
[
Yes
].
2.
Select the desired test items and select [
Start
].
3.
Perform operations following the displayed instructions.
The message “
Test completed
” will be presented when selected tests have been finished. Then you can select
the [
Record
] soft key to record the test result.
NOTE
•
The “Off” position of the Mode Select knob is not tested during the controls test. If you turn the
knob to “Off” for more than 10 seconds, the equipment will be turned off.
•
The tested controls are indicated in green during the controls test.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...