19 - 4
Defibrillator/Monitor Operator’s Manual
CAUTION
•
Use only specified thermal paper. Otherwise, it may cause damage to the recorder’s print head, the
recorder may be unable to print, or poor print quality may result.
•
Never pull the recorder paper with force when a recording is in process. Otherwise, it may cause
damage to the recorder.
•
Do not leave the recorder door open unless you have to reload paper or remove troubles.
19.7
Removing Paper Jam
If the recorder works incorrectly or produces unusual sounds, check if there is a paper jam first. If a paper jam is
detected, follow this procedure to remove it:
1.
Open the recorder door.
2.
Take out the paper and tear off the draped part.
3.
Reload the paper and close the recorder door.
19.8
Cleaning the Recorder Print Head
If the recorder has been used for a long time, deposits of paper debris may collect on the print head
compromising the print quality and shortening the lifetime of the roller. Follow this procedure to clean the print
head:
1.
Take measures against the static electricity, such as disposable wrist strap, for the work.
2.
Open the recorder door and take out the paper.
3.
Gently wipe around the print head using cotton swabs dampened with alcohol.
4.
After the alcohol has completely been dried, reload the paper and close the recorder door.
CAUTION
•
Do not use anything that may destroy the thermal element.
•
Do not add unnecessary force to the thermal head.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...