14 - 4
Defibrillator/Monitor Operator’s Manual
To manually start a zero calibration:
1.
Press the Main Menu button on the front panel, and then select [
Others >>
]
→
[
Maintenance >>
]
→
[
Installation Mode >>
]
→
enter the required password.
2.
Select [
Maintain CO2
]
→
[
Zero
].
Disconnecting the patient airway is not required when performing a zero calibration.
NOTE
•
The CO
2
module temporally stops measuring during zeroing.
14.6
Changing CO
2
Settings
14.6.1
Changing CO
2
Alarm Settings
1.
Select the CO
2
parameter area to enter the [
CO2 Setup
] menu.
2.
Select [
Para. Alarm >>
]
3.
Set the following alarm properties:
◆
Switch on or switch off the alarms or alarm recording.
◆
Adjust the alarm limits or alarm priority.
14.6.2
Setting Pressure Unit
1.
Press the Main Menu button on the front panel, and then select [
Others >>
]
→
[
Configuration >>
]
→
enter the required password.
2.
Select [
CO2 Setup
].
3.
Set [
Press. Unit
].
14.6.3
Changing CO
2
Wave Settings
1.
Select the CO
2
parameter area to enter the [
CO2 Setup
] menu.
2.
Set the wave sweep and scale.
◆
Select [
Sweep
] and then choose an appropriate setting. The faster the wave sweeps, the wider the
wave is.
◆
Select [
Scale
] and adjust the upper scale to change the amplitude of CO
2
waveform.
3.
Select [
Others >>
] and set the wave type:
◆
[
Draw
]: The CO
2
wave is displayed as a curved line.
◆
[
Fill
]: The CO
2
wave is displayed as a filled area.
14.6.4
Changing CO
2
Operating Mode
To change CO
2
operating mode:
1.
Select the CO
2
parameter area to enter the [
CO2 Setup
] menu
2.
Set [
Operating Mode
].
◆
[
Measure
]: only in the Measure mode can you start the CO
2
module.
◆
[
Standby
]: in the standby mode, the gas intake pump, infrared light source, etc. are automatically
switched off to reduce power consumption and extend the lifetime of the CO
2
module.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...