Defibrillator/Monitor Operator’s Manual
24 - 1
24
Maintenance
24.1
Maintenance Introduction
The equipment must be maintained to be ready for immediate use. To ensure proper performance of the
equipment, you should strictly perform the maintenance in this chapter. After the equipment has been used for
12 months, or whenever the equipment is repaired or upgraded, a thorough inspection should be performed to
ensure the reliability.
Make sure to clean and disinfect the equipment before any test and maintenance.
In case of any damage or abnormality, remove the equipment from use. Contact the hospital’s biomedical
engineers or your service personnel immediately.
24.2
Maintenance Safety Information
WARNING
•
Failure for the responsible individual, hospital or institution employing this equipment to
implement a satisfactory maintenance schedule may cause undue equipment failure and possible
health hazards.
•
The safety checks or maintenance involving any disassembly of the equipment should be performed
by professional servicing personnel. Otherwise, undue equipment failure and possible health
hazards could result.
•
If you find a problem with any of the equipment, contact your service personnel or the
manufacturer.
•
No modification of this equipment is allowed.
•
The service personnel must be properly qualified and thoroughly familiar with the operation of the
equipment.
•
Do not open the equipment housings. All servicing and future upgrades must be carried out by the
service personnel.
•
Do not touch, connect or apply any electrode pads and paddles during user test and auto test.
Otherwise, electric shock could result.
24.3
Routine Maintenance
The routine maintenance should be periodically performed in conjunction with the assurance program of the
hospital where the equipment is used. The following test are recommended:
Maintenance Item
Recommend Frequency
Test Item
Shift check
Every shift, or at least every day.
D Defibrillator Shift Checklist
.
Auto test
Automatically, whenever the
equipment is turned on.
Performs tests of the main control board, therapy
module, monitor module, batteries.
Every day, 3:00 am by default
Performs tests of the main control board, therapy
module, monitor module, batteries, 1 J internal
discharge, 10 J internal charge/external discharge
*
.
Every week
Performs tests of the main control board, therapy
module, monitor module, batteries, 1 J internal
discharge, 10 J internal charge/external discharge
*
,
200 J and 360 J internal discharges.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...