Defibrillator/Monitor Operator’s Manual
F - 3
For BF
applied parts
◆
100μA in Normal Condition
◆
500μA in Single Fault Condition
F.7
Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage, through a limiting
resistance, to selected applied part terminals. Current measurements are then taken between the selected
applied part and earth. Measurements are taken with the test voltage (110% of mains) to applied parts in the
normal and reverse polarity conditions
The following outlet conditions apply when performing the Mains on Applied Part test.
■
Normal Polarity
■
Reversed Polarity
LIMITS
■
For CF
applied parts: 50 μA
■
For BF
applied parts: 5000 μA
F.8
Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the remaining
Applied Part connector s. All measurements may have a true RMS only response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
■
normal polarity (Normal Condition)
■
reverse polarity (Normal Condition)
■
normal polarity with open neutral (Single Fault Condition)
■
reverse polarity with open neutral (Single Fault Condition)
■
normal polarity with open earth (Single Fault Condition)
■
reverse polarity with open earth (Single Fault Condition)
LIMITS
For CF
applied parts,
◆
10μA in Normal Condition
◆
50μA in Single Fault Condition
For BF
applied parts,
◆
100μA in Normal Condition
◆
500μA in Single Fault Condition
NOTE
•
Make sure the safety analyzer is authorized comply with requirement of IEC61010-1.
•
Follow the instructions of the analyzer manufacturer.
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...