Defibrillator/Monitor Operator’s Manual
7 - 9
7.7
Contact Impedance Indicator
The contact impedance indicator is used to indicate the impedance between the patient and electrode pads/
external paddles in the Manual Defib mode and AED mode, as shown below:
The display of the contact impedance indicator is switched off by default. To switch it on, set [
Contact
Impedance Indicator
] to [
On
] in the [
Manual Defib Setup
] menu through the Configuration Main menu.
The following table lists the indicator status of the measured impedance and the corresponding actions:
NOTE
•
It is recommended to perform the defibrillation on a patient when the contact impedance indicator
is illuminated in green. If the contact impedance indicator is illuminated in yellow, it also can be
used for the defibrillation. However, the expected effects may not be achieved in this condition.
Contact Impedance
Indicator
Description
Corrective Actions
indicates the patient contact is good, the
impedance is suitable for the defibrillation.
/
(illuminated in green)
indicates the patient contact is not good,
the impedance is slightly higher for the
defibrillation.
Firmly attach the electrode pads or external
paddles to the patient. or adjust the
placements of electrode pads or external
paddles until the indicator is illuminated in
green.
If the indicator is still illuminated in yellow, it
also can be used for the defibrillation. However,
the expected effects may not be achieved in
this condition.
(illuminated in yellow)
indicates the patient contact is very poor,
or there is a short circuit between electrode
pads or external paddles.The impedance is
completely not suitable for the
defibrillation.
Firmly attach the electrode pads or external
paddles to the patient. or adjust the
placements of electrode pads or external
paddles until the indicator is illuminated in
green or yellow.
(illuminated in red)
indicates the ECG cable or therapy cable
falls off.
Check that the ECG cable or therapy cable is
properly connected to the equipment.
(off )
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...