E - 4
Defibrillator/Monitor Operator’s Manual
CO2
CO2 Sensor High Temp
L
C
Check, stop using or replace the sensor.
CO2 No Watertrap
L
B
Check the watertrap connections.
CO2 Zero Failed
L
A
Check the CO
2
connections. After the sensor’s
temperature becomes stabilized, perform a zero
calibration again.
CO2 Module Error
L
C
There is a problem with the CO
2
module, or a problem
with the communications between the
main unit
and
the CO
2
module. Restart the equipment.
CO2 Occlusion
L
C
The airway was occluded. Check the airway and remove
the occlusion.
CPR sensor
CPR Sensor Err
H
C
There is a self-test error or communication problem with
the CPR sensor. Contact your service personnel.
CPR Sensor Low Battery
M
C
The battery power of the CPR sensor is low. Charge the
battery by connect the CPR sensor to the equipment.
CPR Sensor Need
Service
H
C
The compressions using the CPR sensor exceed the
expected numbers. Contact your service personnel.
CPR Sensor Cable Fault
L
C
An error occurred to the CPR sensor cable. Replace the
CPR sensor cable.
Change CPR Sensor
Battery
L
C
The CPR sensor battery is aging. Contact your service
personnel.
CPR Sensor Bat. Charge
Err
L
C
The CPR sensor cannot be charged. Contact your service
personnel.
Main control
system
No Speaker
L
C
Make sure that the speaker is connected.
Power Board Comm Err
H
C
An error occurred to the power board, or there is a
problem with the communications between the power
board and the
main unit
. Restart the equipment.
Keyboard Comm Err
L
C
An error occurred to the keypad board, or there is a
problem with the communications between the keypad
board and the
main unit
. Restart the equipment.
Therapy Module Comm
Err
S
C
An error occurred to the therapy module, or there is a
problem with the communications between the therapy
module and the
main unit
. Restart the equipment. If the
problem persists, contact your service personnel.
Main Control Selftest
Err
H
C
The main control voltage is abnormal. Replace the main
control board.
Wifi Module Fault
L
C
Contact your service personnel.
Machine Type Error
H
C
RT Clock Need Reset
L
C
Reset system time.
RT Clock Err
H
C
An error occurred to the RTC chip, or the button cell is
depleted. Replace corresponding part.
Memory Error
L
C
There is a problem with the data card. Format the CF card.
If the problem persists, contact your service personnel.
Last User Test Failed
L
C
Run a successful user test.
Last Auto Test Failed
L
C
Run a successful user test again.
NO CMS.
L
C
The equipment is disconnected from the CMS. Check the
network connection.
IP Address Conflict
L
C
Network IP conflicts. Check the network settings.
Measurement
Alarm Message
L
I
Cause and solution
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...