Operator's Manual
WP700Zi-OM-E-RevA
80
2.
Choose a function, and select
ExcelMath
as
Operator1
for the
function. Excel will open automatically in the background.
3.
Choose File, Minimize from the menu bar to minimize the instrument display and open the Excel
program.
4.
Create your formula for each data point in column A (in this case, our formula for cell C2 is
–A2
,
copied for the entire column), as shown
here
.
5.
Retrigger the scope (if it is not currently triggering)
6.
Return to the program
Note that the only action that was needed in Excel was to create the formula in column C for each data point in
column A. The instrument automatically opens Excel, puts the waveform data in the correct columns, and returns
the calculated data back to the display as the chosen F trace. This Excel-calculated trace can have further
measurements or math calculations performed on it, if desired.