Operator's Manual
WP700Zi-OM-E-RevA
442
For example, the FSB mode requires setting up multiple channels to carry out the eye diagram measurements,
as described in the example below. The Gated Eye Diagram requires at least two inputs: The Data and the Gate
signal. The Transition Mode requires at least one input (the Data Stream, but depending on the specific serial
data standard, a clock input may also be required to recover the clock information).
These signals (except for the strobes) are differential signals, so the use of differential probes is often required to
carry out the measurements. As with any high-speed precision measurements, all probes and SMA cables should
be calibrated and deskewed prior to any data collection.
Example Setups
FSB Eye Mode Configuration
Channel Setup
x
DataStrobep: Channel 1 (M1)
x
DataStroben: Channel 2 (M2)
x
Data1: Channel 3 (M3)
x
Data2: Channel 4 (M4)
Serial Data Analysis Setup
1.
Select Signal Type: FSB 533MHz, 667MHz or 800MHz
2.
Uncheck Recover Clock -- No clock recovery is required in FSB case
3.
Select Eye Diagram Tab, then configure the screen as follows:
Mode:
FSB
Data Source 2: Channel 4 (M4)
DSTBP: Channel 1 (M1)
DSTBN: Channel 2 (M2)
The screen should look like this.
Summary of Contents for DDA 7 Zi series
Page 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Page 2: ... L R R H HUD RU D D ...
Page 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Page 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Page 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Page 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Page 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...