WavePro 7Zi
267
WP700Zi-OM-E-RevA
Button: Description:
Trigger Out
— can be used to trigger another scope.
DC level —
a reference level
Trigger Enabled
— can be used as a gating function to trigger another instrument when the scope is
ready
Pass/Fail
— allows you to set a pulse duration from 1 ms to 500 ms; generates a pulse when
pass/fail testing is active and conditions are met.
Aux Output Off
— turns off the auxiliary output signal
Setting Up Auxiliary Output
1. Touch
Utilities Utilities Setup
on the menu bar.
2. Touch
the
Aux Output
tab.
3.
If you want a 1 kHz, 1 V Square Wave, touch the
Set to 1 kHz, 1 V Square Wave
button.
OR
If you want a specialized output, touch one of the following buttons under
Use Auxiliary Output For
:
Square Wave
,
Trigger Enabled
,
Trigger Out
,
Pass/Fail
,
or Off
.
Note
: If you choose
Pass/Fail
, touch inside the
Pulse Duration
field (if applicable) and enter a value from 1 ms
to 500 ms, using the pop-up numeric keypad.
4.
If applicable, touch inside the
Amplitude
data entry field and enter a value, using the slider bar at the
bottom of the window. Click the keypad button
on the slider bar to enter a value using the pop-
up numeric keypad.
5. Touch
inside
the
Frequency
data entry field and enter a value, using the slider bar at the bottom of the
window. Click the keypad button
on the slider bar to enter a value using the pop-up numeric
keypad.
Date and Time
From the
Date and Time
tab on the
Utilities
dialog, you can manually set the time and date or get it from the
Internet. If you elect to get the date and time from the Internet, you need to have the oscilloscope connected to
the Internet through the LAN connector on the side panel. You can also set time zones and daylight savings time.
Setting the Date and Time Manually
1. Touch
Utilities Utilities Setup
on the menu bar.
2. Touch
the
Date/Time
tab.
3.
Touch inside each of the
Hour
,
Minute
,
Second
,
Day
,
Month
, and
Year
data entry fields and enter a
value, using the pop-up numeric keypad.
4. Touch
the
Validate Changes
button.
Setting the Date and Time from the Internet
Note
: The Simple Network Time Protocol (SNTP) is used.
1.
Ensure that the scope is connected to the Internet through the LAN connector at the side of the scope.
2. Touch
Utilities Utilities Setup
on the menu bar.
Summary of Contents for DDA 7 Zi series
Page 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Page 2: ... L R R H HUD RU D D ...
Page 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Page 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Page 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Page 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Page 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...