Operator's Manual
WP700Zi-OM-E-RevA
520
Triggering System
WavePro 715Zi
SDA, WavePro
725Zi
SDA, DDA,
WavePro
735Zi
SDA, WavePro
740Zi
SDA, DDA,
WavePro 760Zi
Edge Trigger (Ch 1-4)
ProLink Inputs
1.5 div @ 2 GHz
1.0 div @ 200
MHz
(for DC, AC,
LFRej coupling,
10 mV/div, 50 )
1.5 div @ 3 GHz
1.0 div @ 200
MHz
(for DC, AC,
LFRej coupling,
10 mV/div, 50 )
External Trigger
Sensitivity, (Edge
Trigger)
2 div @ < 1 GHz
1.5 div @ < 500 MHz
1.0 div @ < 200 MHz
(for DC, AC, LFRej coupling)
Max, Trigger
Frequency, SMART
Trigger™
1.0 GHz @ 10
mV/div
(minimum
triggerable width
500 ps)
2.0 GHz @ 10
mV/div
(minimum
triggerable width
500 ps)
2.0 GHz @
10 mV/div
(minimum
triggerable
width 500 ps)
2.0 GHz @ 10 mV/div
(minimum triggerable width 500 ps)
External Trigger Input
Range
Aux (±0.4 V); Aux/10 (±4 V)
Basic Triggers
Edge
Triggers when signal meets slope (positive, negative, or either) and level
condition.
TV-Composite Video
Triggers NTSC or PAL with selectable line and field; HDTV (720p, 1080i,
1080p) with selectable frame rate (50 or 60 Hz) and Line; or CUSTOM with
selectable Fields (1-8), Lines (up to 2000), Frame Rates (25, 30, 50, or 60
Hz), Interlacing (1:1, 2:1, 4:1, 8:1), or Synch Pulse Slope (Positive or
Negative).
Window
Trigger when signal or exits a window defined by adjustable thresholds.
SMART Triggers
State or Edge Qualified
Triggers on any input source only if a defined state or edge occurred on
another input source.
Delay between source is selectable by time or events.
Qualified First
In Sequence acquisition mode, triggers repeatably on event B only if a
defined pattern, state, or edge (event A) is satisfied in the first segment of the
acquisition. Delay between sources is selectable by time or events.
Dropout
Triggers if signal drops out for longer than selected time between 1 ns and
20 s.
Pattern
Logic combination (AND, NAND, OR, NOR) of 5 inputs (4 channels and
external trigger input).
Each source can be high, low, or don't care. The High and Low level can be
selected independently.
Triggers at start or end of the pattern.
SMART Triggers with Exclusion Technology
Glitch
Triggers on positive or negative glitches with widths selectable as low as 200
ps (depending on oscilloscope bandwidth) to 20 s, or on intermittent faults.
Width (Signal or Pattern)
Triggers on positive, negative, or both widths with widths selectable as low
as 200 ps (depending on oscilloscope bandwidth) to 20 s, or on intermittent
faults.
Interval (Signal or Pattern)
Triggers on intervals selectable between 1 ns and 20 s.
Timeout (State/Edge Qualified)
Triggers on any source if a given state (or transition edge) has occurred on
another source.
Summary of Contents for DDA 7 Zi series
Page 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Page 2: ... L R R H HUD RU D D ...
Page 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Page 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Page 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Page 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Page 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...