Operator's Manual
WP700Zi-OM-E-RevA
404
Tj values at BER levels below 10
-10
. The jitter breakdown is the best-fit to the bathtub curve for very low BER
values, but does not take into account the jitter contribution at the very top of the bathtub curve. The figure below
shows the flow of the effective jitter measurement.
EffeRj
Effective Random Jitter is one of the two constants used to estimate the function Tj(BER)
for very low BER. Rje is obtained through the analysis of the growth of the extrapolated histogram
of TIE as a function of BER, fitting the form: Tj(BER) = Dje + Tj
(sigma=1)
(BER)*Rje
EffeDj
Effective Deterministic Jitter is one of the two constants used to estimate the function
Tj(BER) for very low BER. Rje is obtained through the analysis of the growth of the extrapolated
histogram of TIE as a function of BER, fitting the form: Tj(BER) = Dje + Tj
(sigma=1)
(BER)*Rje
MJSQ Jitter
The other method of determining the random and deterministic components is based on the procedure described
in the Fibrechannel MJSQ (Methods for Jitter and Signal Quality) document. This method is similar to the effective
jitter described above except that it operates directly on the jitter probability density function (pdf). A pair of normal
distributions (one for each side) is found by adjusting their standard deviations so that they best fit the tails of the
measured pdf. The difference between their mean values is the Dj and the average of their standard deviations is
the Rj. This method is shown below.
Summary of Contents for DDA 7 Zi series
Page 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Page 2: ... L R R H HUD RU D D ...
Page 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Page 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Page 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Page 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Page 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...