Scan Filters
But, now we want to look at fall times greater than 3 ns; fifteen falling edges from the same acquisition pass this
new filter:
You can also decide how anomalous the features located should be by using the
Filter Wizard
to select the
average value as the limit, to search for rarest events, or to apply statistical criteria.
Filter Wizard
You can decide how anomalous the features searched or scanned for should be by using the average value as
the limit, by searching for rarest events, or by applying statistical criteria: ±1, 3, or 5 sigma.