WavePro 7Zi
295
WP700Zi-OM-E-RevA
Figure 5-6. Zoom of Positive Polarity Pit Ending Edge -- example
The measurement has configurable units. If absolute time is specified, the value is simply the time as indicated
above. If percent is specified, the value of the measurement is the time normalized to the clock period:
For all pits, a valid measurement will be obtained only when both pit/space edges can be determined (that is,
there is a hysteresis-qualified threshold crossing beginning and ending the pit/space pair of interest between the
parameter cursors), and there is a clock edge of both polarities surrounding the ending pit or space edge between
the parameter cursors.
Display Options
ORM parameter calculations can be displayed, histogrammed, and trended in a variety of ways.
DISPLAY TYPE
VALUE DISPLAYED
Parameter
Statistics Off
All values of the average time between ending edge of the subject n pit (space) and nearest
clock edges for all subject pits (spaces) followed by the spaces (pits) within the selected ‘nT’
range for the last acquisition.
Parameter
Statistics On
Average, minimum, maximum, and sigma of the ending edge shift calculated for all
identified pits/spaces pairs within the selected ‘nT’ range for all acquisitions since the last
CLEAR SWEEPS operation.
nT Table
List of values of the average ending edge shift for each ‘nT’ space (pit) within the selected
range following the subject pit (space) for the last acquisition.
Histogram
Function
Histogram graph of the value of the ending edge shift calculated for all pit/space pairs within
the selected ‘nT’ range for all acquisitions since the last CLEAR SWEEPS operation.
Trend Function
Trend graph of the value of the ending edge shift calculated for all pit/space pairs within the
selected ‘nT’ range for all acquisitions since the last CLEAR SWEEPS operation.
Ending Edge Shift Sigma (EESS)
EESS provides a measurement of the mean, normalized standard deviation of the Ending Edge Shift
measurements (see
EES
). When a single
n
is specified, or when you are in ‘nT Table’ Show mode, the value
calculated for the nth index is calculated using the following equation for standard deviation:
Summary of Contents for DDA 7 Zi series
Page 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Page 2: ... L R R H HUD RU D D ...
Page 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Page 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Page 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Page 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Page 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...