Operator's Manual
WP700Zi-OM-E-RevA
256
4.
If available, touch inside the
Coupling
field and select a
coupling
mode.
5.
Select
Positive
or
Negative
polarity.
6. Touch
inside
the
Trigger after timeout if NO edge within
data entry field and enter a time value using the
slider bar at the bottom of the window. Click the keypad button
on the slider bar to enter a value
using the pop-up numeric keypad. Touch the Default button
to use a preset value.
7.
Enter a voltage level in the
Level
field using the slider bar at the bottom of the window. Click the keypad
button
on the slider bar to enter a value using the pop-up numeric keypad. Touch the Default
button
to use a preset value. The Find Level button
is available to
automatically set the Level.
Note
: Use the
TriggerScan
tab to run the
Trainer
(a wizard-like tool to configure triggers),
Save
, and
Load
trigger setups for reuse.
Runt Trigger
Not available on all instruments, the
Runt
trigger is programmed to occur when a pulse crosses a first threshold
line and fails to cross a second threshold line before re-crossing the first. You can select both voltage thresholds
within a time range of 100 ps to 20 s. Other defining conditions for this trigger are the edge (triggers on the slope
opposite to that selected) and runt width.
Runt Trigger Setup
1. Touch
the
Trigger
descriptor label.
2.
If applicable, touch the
Smart
trigger button,
and then touch the
Runt
trigger button.
3.
Touch inside the trigger
Source
field and select a source
on which to trigger. If
you select an external trigger source, you also need to set a
coupling
mode.
Summary of Contents for DDA 7 Zi series
Page 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Page 2: ... L R R H HUD RU D D ...
Page 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Page 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Page 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Page 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Page 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...